Quantification of velocity reduction after flow diverter placement in intracranial aneurysm: An ex vivo study with 3D printed replicas Academic Article uri icon

Overview

MeSH Major

  • Intracranial Aneurysm
  • Printing, Three-Dimensional
  • Replica Techniques

abstract

  • Phase contrast MRI (pcMRI) was used to measure flow before and after placement of a flow diverter (n = 3). Decreases from 18% to 31% in flow velocity were seen in the inflow jet of the aneurysms. Flow patterns were also compared. It was observed that the gross aneurysmal flow patterns were maintained after flow diverter placement despite decreased fluid velocities. All measurements were carried out in 3D printed aneurysm replicas.

publication date

  • November 4, 2015

Research

keywords

  • Academic Article

Identity

Language

  • eng

Digital Object Identifier (DOI)

  • 10.1109/EMBC.2015.7320077

PubMed ID

  • 26737977

Additional Document Info

start page

  • 7300

end page

  • 3

volume

  • 2015-November