Validation of multiple whole slide imaging scanners based on the guideline from the College of American Pathologists pathology and laboratory quality center Academic Article uri icon

Overview

MeSH Major

  • Diagnostic Imaging
  • Image Interpretation, Computer-Assisted
  • Neoplasms
  • Pathology, Clinical
  • Pathology, Surgical

abstract

  • The results were felt to validate the use of WSI for the intended applications in our multiinstitutional laboratory system, although scans at ×20 magnification may be insufficient for cases hinging on small focal findings, such as microorganisms and inflammatory processes.

publication date

  • January 2015

Research

keywords

  • Academic Article

Identity

Language

  • eng

Digital Object Identifier (DOI)

  • 10.5858/arpa.2014-0073-OA

PubMed ID

  • 25927149

Additional Document Info

start page

  • 656

end page

  • 64

volume

  • 139

number

  • 5