Comparison of conforming and nonconforming retrieved glenoid components. Academic Article Article uri icon

Overview

MeSH

  • Aged
  • Arthroplasty, Replacement
  • Female
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Failure
  • Reoperation
  • Retrospective Studies

MeSH Major

  • Joint Prosthesis
  • Prosthesis Design
  • Shoulder Joint

abstract

  • The purpose of this study was to compare differences in wear performance of conforming and nonconforming glenoid designs, using clinical, radiographic, and retrieved polyethylene glenoid component outcome Sixty-three glenoids met the study criteria, and each glenoid was assigned to the conforming group (if the radii of curvature of the humeral and glenoid components were identical) or the nonconforming group (if a mismatch existed between the radii of curvature). The mean length of implantation was 5.6 +/- 5.5 years in the conforming group and 3.1 +/- 3.1 years for the nonconforming group (P < .05). The loosening score was 3.2 +/- 2.0 in the conforming group and 2.4 +/- 1.2 in the nonconforming one (P < .05). The nonconforming group had a significantly greater burnishing score (P < .01), while the conforming group had greater abrasion and delamination scores (P < .05). Articular conformity contributes to differences observed from retrieved polyethylene glenoid components, which are consistent with differences in performance that may influence loosening.

publication date

  • November 2008
  • December 2008

has subject area

  • Aged
  • Arthroplasty, Replacement
  • Female
  • Humans
  • Joint Prosthesis
  • Male
  • Middle Aged
  • Prosthesis Design
  • Prosthesis Failure
  • Reoperation
  • Retrospective Studies
  • Shoulder Joint

Research

keywords

  • Comparative Study
  • Journal Article

Identity

Language

  • eng

Digital Object Identifier (DOI)

  • 10.1016/j.jse.2008.04.010

PubMed ID

  • 18799324

Additional Document Info

start page

  • 914

end page

  • 920

volume

  • 17

number