Development of Neuromonitoring Pedicle Screw - Results of Electrical Resistance and Neurophysiologic Test in Pig Model. Academic Article uri icon

Overview

abstract

  • OBJECTIVE: To analyze the electrical resistance of a newly developed neuromonitoring pedicle screw (Neuro-PS) and to verify the electrophysiologic properties of the Neuro-PS in a pig model. METHODS: We developed 2 types of the Neuro-PS in which a gold lead was located internally (type I) and externally (type II). We measured the electrical resistance of the Neuro-PS and the conventional screw and analyzed the electrical thresholds of triggered EMG (t-EMG) of each screw by intentionally penetrating the medial pedicle wall and contacting the exiting nerve root in a pig model. RESULTS: The electrical resistances of the Neuro-PS were remarkably lower than that of the conventional screw. In electrophysiologic testing, only the type II Neuro-PS under the leadnerve contact condition showed a significantly lower stimulation threshold as compared to the conventional screw. CONCLUSION: The Neuro-PS demonstrated lower electrical resistances than the conventional screw. The type II Neuro-PS under the lead-nerve contact condition showed a significantly lower stimulation threshold compared to that of the other screws in the t-EMG test.

publication date

  • November 18, 2020

Identity

PubMed Central ID

  • PMC8021819

Scopus Document Identifier

  • 85114120261

Digital Object Identifier (DOI)

  • 10.14245/ns.2040424.212

PubMed ID

  • 33211943

Additional Document Info

volume

  • 18

issue

  • 1