  • Performance of a second itG 68Ge/68Ga generator system and production of 68Ga-DOTATOC and 68Ga-PSMA-HBED-CC were tested over one year as an accompaniment to a previously published study (J Nucl Med. 2016;57:1402-1405). METHODS: Performance of a 1951MBq 68Ge/68Ga generator was characterized and the eluate used for preparation of 68Ga-DOTATOC and 68Ga-PSMA-HBED-CC. Weekly elution profiles of 68Ga elution yield and 68Ge breakthrough were determined. RESULTS: 68Ga elution yields averaged 82% (61.8-98.4%) and 68Ge breakthrough averaged 0.002% (0.0007% to 0.004%). The radiochemical purities of 68Ga-DOTATOC and 68Ga-PSMA-HBED-CC were determined by HPLC analysis to be >98% and specific activity was 12.6 and 42GBq/μmol, respectively. 68Ge contamination in the product was under the detection limit (0.00001%). Final sterile, pyrogen-free formulation of 68Ga-DOTATOC and 68Ga-PSMA-HBED-CC in physiologic saline with 5%-7% ethanol was achieved. CONCLUSION: Performance of a 68Ge/68Ga generator was studied over one year with satisfactory results. The generator eluate was used to synthesize 68Ga-DOTATOC and 68Ga-PSMA-HBED-CC on a routine basis in high purity.