Comprehensive Quality Control of the ITG 68Ge/68Ga Generator and Synthesis of 68Ga-DOTATOC and 68Ga-PSMA-HBED-CC for Clinical Imaging. Academic Article uri icon

Overview

abstract

  • UNLABELLED: A good-manufacturing-practices (GMP) (68)Ge/(68)Ga generator that uses modified dodecyl-3,4,5-trihydroxybenzoate hydrophobically bound to a octadecyl silica resin (C-18) as an adsorbent has been developed that allows for dilute HCl (0.05N) to efficiently elute metal-impurity-free (68)Ga(3+) ready for peptide labeling. We characterized the performance of this generator system over a year in conjunction with the production of (68)Ga-labeled DOTATOC and Glu-NH-CO-NH-Lys(Ahx)-HBED-CC (PSMA-HBED-CC) intended for clinical studies and established protocols for batch release. METHODS: A 2,040-MBq self-shielded (68)Ge/(68)Ga generator provided metal-free (68)GaCl3 ready for peptide labeling in the fluidic labeling module after elution with 4 mL of 0.05N HCl. The compact system was readily housed in a laminar flow cabinet allowing an ISO class-5 environment. (68)Ga labeling of peptides using GMP kits was performed in 15-20 min, and the total production time was 45-50 min. Batch release quality control specifications were established to meet investigational new drug submission and institutional review board approval standards. RESULTS: Over a period of 12 mo, (68)Ga elution yields from the generator averaged 80% (range, 72.0%-95.1%), and (68)Ge breakthrough was less than 0.006%, initially decreasing with time to 0.001% (expressed as percentage of (68)Ge activity present in the generator at the time of elution), a unique characteristic of this generator. The radiochemical purity of both (68)Ga-DOTATOC and (68)Ga-PSMA-HBED-CC determined by high-performance liquid chromatography analysis was greater than 98%, with a minimum specific activity of 12.6 and 42 GBq/μmol, respectively. The radionuclidic ((68)Ge) impurity was 0.00001% or less (under the detection limit). Final sterile, pyrogen-free formulation was provided in physiologic saline with 5%-7% ethanol. CONCLUSION: The GMP-certified (68)Ge/(68)Ga generator system was studied for a year. The generator system is contained within the fluidic labeling module, and it is compact, self-shielded, and easy to operate using simple manual techniques. The system provides radiolabeled peptides with high (>98%) radiochemical purity and greater than 80% radiochemical yield. The (68)Ge levels in the final drug products were under the detection limits at all times. (68)Ga-DOTATOC and (68)Ga-PSMA-HBED-CC investigational radiopharmaceuticals are currently being studied clinically under investigational new drug (IND) applications submitted to the U.S. Food and Drug Administration.

publication date

  • April 21, 2016

Research

keywords

  • Edetic Acid
  • Octreotide
  • Oligopeptides
  • Organometallic Compounds
  • Radionuclide Generators

Identity

Scopus Document Identifier

  • 84989282806

Digital Object Identifier (DOI)

  • 10.2967/jnumed.115.171249

PubMed ID

  • 27103024

Additional Document Info

volume

  • 57

issue

  • 9