LGM vena cava filter: objective evaluation of early results. Academic Article uri icon

Overview

abstract

  • One hundred one LG-Medical (LGM) vena cava filters were placed in 97 patients at four institutions. Placement was a complete technical success in 90% (91 of 101). In 6% of attempts, LGM filter insertion was complicated by incomplete opening of the filter. Pulmonary embolism after filter placement was not definitely demonstrated in any patient. The probability of inferior vena cava patency was 92% at 6 months after filter insertion. Thrombosis at the insertion site was seen in eight of 35 patients (23%) evaluated with duplex ultrasound or venography. Thrombus was observed in 37% of filters at follow-up examination, with cephalic extension of thrombus above the filter in 20% of all patients examined. Filter migration (greater than 1 cm) was seen in 12%; significant angulation was observed in only one patient (2%). In vitro experimentation demonstrated that incomplete opening of the LGM filter during placement can be avoided, in part, by brisk retraction of the insertion cannula. The low-profile introducer system of the LGM filter allows increased alternatives in selecting the site for filter insertion. The low-profile system also makes outpatient filter placement a possibility. No significant difference in the prevalence of thrombosis at the insertion site following LGM filter insertion was noted compared with previous results reported for percutaneous transfemoral placement of the Greenfield filter. The nonopaque sheath does not permit careful localization prior to filter deposition. Modification of the LGM filter to include a radiopaque sheath is suggested.

publication date

  • February 1, 1991

Research

keywords

  • Pulmonary Embolism
  • Thrombosis
  • Vena Cava Filters

Identity

Scopus Document Identifier

  • 0026102445

PubMed ID

  • 1799739

Additional Document Info

volume

  • 2

issue

  • 1