Discrepancies between the upper limit of vulnerability and defibrillation threshold: Prevalence and clinical predictors Academic Article uri icon

Overview

MeSH Major

  • Defibrillators, Implantable
  • Electric Countershock
  • Equipment Failure Analysis
  • Tachycardia
  • Ventricular Fibrillation

abstract

  • Significant discrepancies between ULV and DFT occur in 5% of patients with ICDs. Absence of coronary disease and documented ventricular arrhythmias predict such a discrepancy. At ICD implant, DFT testing is recommended in these patients and in patients with a high (>20 J) ULV before first-shock energy and the need for lead repositioning are determined.